Cat plague virus antigen (FPV Ag) test kit is used for rapid qualitative detection of feline pestilential virus antigen in cat feces, and can be used for screening and auxiliary diagnosis of feline pestilential virus infection.MOQ:500
Cat plague virus antigen (FPV Ag) test kitFeline Pestilential virus antigen (FPV Ag) detection kit is used for rapid qualitative detection of feline pestilential virus antigen in cat feces, and can be used for screening and auxiliary diagnosis of feline pestilential virus infection.
This kit uses double antibody sandwich immunochromatography. If the sample contains a sufficient amount of cat plague.The virus antigen, the cat plague virus antigen binds to the monoclonal antibody coated in colloidal gold on the gold tag,Forming an antibody-antigen complex. This complex migrates upward to the detection line (T-line) with the capillary effect.It then binds to another monoclonal antibody to form an "antibody-antigen-antibody" complex and gradually agglutinates.Forming a visible detection line (T-line), excess colloidal gold antibodies continue to migrate to the quality control line (C-line)
This kit is stored at 2-30℃; Do not freeze. Valid for 24 months; test kit After opening the bag, use the reagent as soon as possible.
1. Test sample: cat feces
2. Samples should be tested on the same day; Samples that cannot be tested on the same day should be stored at 2-8 ° C over For 24 hours, it should be stored at -20℃.
1. Before use, restore the kit to room temperature (15-30℃).
2. Remove the reagent card from the foil bag and place it on a clean platform.
3. Unscrew the top tube cap on the diluent tube cover containing the sample, invert the diluent tube, and squeeze
The wall of the tube adds 3-5 drops of the sample mixture into the sample hole (S hole) of the reagent card.
4. The results can be read in 10-15 minutes. The result is invalid after 15 minutes.
Positive: Both the quality control line (C line) and the test line (T line) appear
Negative: Only the quality control line (C line) is available
Invalid: Quality control line does not appear, take a new device to retest